APPENDIX C - CIIL COMMAND DEFINITIONS
C.1 INTRODUCTION
This appendix defines the CIIL commands used with the MST 488-27 Controller. Table C-1 pro-
vides a quick reference of all CIIL commands used in the MST 488-27.
FIGURE C-1. FNC — FUNCTION COMMAND
TABLE C-1. CIIL SUBSYSTEM COMMAND/QUERY INDEX
COMMAND PAGE COMMAND PAGE
CLS C-4 OPN C-4
CNF C-4 RST C-4
FNC C-1 SET C-3
FTH C-2 SRN C-3
GAL C-6 SRX C-3
INX C-2 STA C-5
IST C-4
Syntax: Stimulus mode: FNC DCS :CHnn <SET Command>
Sensor mode: FNC DCS <VOLT or CURR command> :CHnn
Function: This operator is used with either the SET command to program a power supply's output (stimulus
mode), or with the VOLT and CURR commands to read its output settings (sensor mode).
Description: The first operand contains the three (3) letter mnemonic pertaining to the device on the control bus, in
this case DCS (Direct Current Source). If a reading is being set up, the modifier VOLT or CURR fol-
lows. The next operand is used to select the specific channel of the device being programmed or read
from. The MST 488-27 can control up to 27 power supplies/power modules with control bus addresses
in the range of 1 to 31.
Example: FNC DCS :CH12 SET VOLT 15 Power supply at node address 12 commanded to 15V
FNC DCS :CH12 SET CURR 3 Power supply at node address 12 commanded to 3A
FNC DCS VOLT :CH03 Power supply at node address 3 returns value which represents
actual output voltage
FNC DCS CURR :CH21 Power supply at node address 21 returns value which represents
actual output current
NOTE: Actual output voltage and current depends on whether output is enabled or disabled
and load conditions